DetAPCn hydrogen set point not reached [Error Code:2015 2018 2021 2024]
Cause¶
The pressure or flow rate may have not reached the set value due to the following reasons.
- The gas supply pressure is insufficient.
 - There is a leak at tubing connections.
 - The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
 
Solution¶
- 
Check that the gas main valve is open.
 - 
Check that the gas supply pressure is within the range below.
 - 
Check that the supply pressure is sufficient for the hydrogen gas to flow.
 - 
Check that no tubing connections are loose and there is no leak.
 - 
Contact your Shimadzu service representative to check the inner diameter and length of used tubing.