DAMCn makeup set point not reached [Error Code:2272 2273 2274 2275]
Cause¶
The pressure or flow rate may have not reached the set value due to the following reasons.
- The gas supply pressure is insufficient.
 - There is a leak at tubing connections.
 - The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
 
Solution¶
- Check that the gas main valve is open.
 - Check that the gas supply pressure is within the range below.
 - Check that the supply pressure is sufficient for the makeup gas to flow.
 - Check that no tubing connections are loose and there is no leak.
 - Contact your Shimadzu service representative to check the inner diameter and length of used tubing.