Not execute this operation [Error Code:9061]
Cause¶
The following items may be the cause. Check the following items.
- Operations such as Ignition/illumination were executed while the GC stopped.
 - Operations such as standard diagnosis were executed while the GC started.
 - Time program setting was changed while time program was running.
 - Pre-Run program setting was changed while pre-run program was running.
 - AOC control was attempted when the instrument and AOC were not connected.
 - Calibration was performed in a situation where calibration was not possible.
 
Solution¶
Operations such as Ignition/illumination were executed while the GC stopped.¶
The following operation was executed while the GC stopped.
- Ignition/Illumination
 - FTD Feedback
 - Start analysis
 - Batch
 - Clean Up
 - PREP RUN
 - Inj Maintenance
 
Press [Start GC] on [GC Start Sequence] screen before executing the listed operations.
Operations such as standard diagnosis were executed while the GC started.¶
The following operation was executed while the GC started.
- Standard Diagnosis
 - Carrier gas leak check
 - Offset calibration of flow controller
 
Press [Stop GC] on [GC Stop Sequence] screen before executing the listed operations.
Time program setting was changed while time program was running.¶
Change the setting after time program is completed.
Pre-Run program setting was changed while pre-run program was running.¶
Change the setting after pre-run program is completed.
AOC control was attempted when the instrument and AOC were not connected.¶
Check the following items.
- 
Connect the instrument and AOC.
 - 
AOC is installed correctly.
Installation_AOC-30i
Installation_AOC-20i - 
AOC power is [On].
 
Calibration was performed in a situation where calibration was not possible.¶
Check the following items before calibration.
- Stopping GC
 - Turn Off FLOW.
 - Open each unit to the atmosphere, when the pressure higher than the specified value has already been applied.